Publishing House SB RAS:

Publishing House SB RAS:

Address of the Publishing House SB RAS:
Morskoy pr. 2, 630090 Novosibirsk, Russia



Advanced Search

Avtometriya

2018 year, number 4

METHOD OF CALIBRATION OF A LASER-BASED GALVANOMETRIC SCANNING SYSTEM WITH SUBMICRON RESOLUTION

V. P. Bessmeltsev, N. V. Goloshevskii, V. V. Kasterov, Ya. A. Kipriyanov, K. K. Smirnov
Institute of Automation and Electrometry, Siberian Branch, Russian Academy of Sciences, 630090, Novosibirsk, prosp. Akademika Koptyuga, 1
Keywords: лазерная микрообработка, гальванометрические дефлекторы, комплиментарные системы сканирования, метод кросскорреляции, laser micromachining, galvanometric deflectors, complementary scanning systems, cross-correlation method

Abstract

A method used for calibration of scanning modules on the basis of galvanometric deflectors is considered. The method is based on an iterative two-step algorithm of processing of the profilometry map of the test recording of the calibrated module. It is demonstrated that high accuracy and robustness of determining the coordinates of the reference points of the test recording can be obtained by means of finding the approximate positions of such points and identifying the fragment in the vicinity of these points at the first step and searching for the exact values of the coordinates by the correlation method at the second step. The use of this method and a simple algorithm of piecewise-planar interpolation offers a possibility of laser beam positioning with a relative error of 10-5 from the recording field.