Publishing House SB RAS:

Publishing House SB RAS:

Address of the Publishing House SB RAS:
Morskoy pr. 2, 630090 Novosibirsk, Russia



Advanced Search

Thermophysics and Aeromechanics

2025 year, number 2

Determination of the amplitude-frequency characteristics of the system consisting of a hot-wire anemometer and a sensor on the basis of the laser pulse: methodology improvement

D.A. Bountin, O.I. Vishnyakov
Khristianovich Institute of Theoretical and Applied Mechanics SB RAS, Novosibirsk, Russia
Keywords: constant-temperature anemometer, amplitude-frequency characteristics, wire sensor, pulsed action

Abstract

We propose to improve a new experimental methodology of determining the amplitude-frequency characteristics of a system of hot-wire anemometry with the use of a short-pulse laser action on the hot-wire sensor. Mistakes of the previous study are taken into account. The amplitude-frequency characteristics of a constant-temperature anemometer with a wire sensor are obtained, and comparisons with the results obtained by a standard method of determining the amplitude-frequency characteristics are performed.